5.
Select the plate records to add, then click Add.
6.
When the needed plate records are listed, click Done.
7.
As needed, click Up, Down, or Remove to organize the list of plate records in the
Input Stack list.
8.
Load the plates in the In Stack in the order to match the corresponding plate
record. The boom plate runs rst and uses the rst plate record in the list.
IMPORTANT! The order of the plate records must match the stack order of the
plates in the In Stack. If the order does not match, data les will contain the
wrong plate record information.
Note: Y
ou can assign more plates in the Run Scheduler than are actually
available in the stacker.
9.
Click
(Run).
During a manual run, the following steps occur
.
• Plates are listed in the Input Stack
• The currently running plate is listed in the Auto Sampler eld
• Completed plates are listed in the Output Stack
For more information, see “Monitor the status of the run“ on page 65.
Change injection order of plate quadrants
You can change the run order of quadrants. Changing the order of the quadrants
overrides the seing in the plate record and is applied to all plates.
1.
Select
GA Instruments4 ga37304Instrument name4 Run Scheduler.
2.
At the top of the screen, select Instrument4 Scheduling Preference.
Not
e: The Instrument menu is available only when the Run Scheduler is
displayed.
3.
Select the quadrant priority (run order)
from the Quadrant list.
You can select any run order. For
information on the order of well injection,
see the following sections.
• “Default injection positions—48-
capillary arrays“ on page 37
• “Default injection positions—96-
capillary arrays“ on page 39
• “Default injection order on plates“ on page 59
Chapter 5 Run the instrument
Change injection order of plate quadrants
5
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
63