Place the Water and Waste reservoirs into the instrument. The reservoirs must be
in the following order from left to right:
1
Buffer reservoir
2
Water reservoir
3
Waste reservoir
4.
Close the instrument door
.
Fill the anode buffer jar
Replace the anode buer at the following times.
• Before each group of scheduled runs, or at least every 48 hours
• After you replenish polymer
• After you replace the capillary array
• After you run the Bubble Remove wizard
• Every time that you change the buer reservoir
IMPORTANT! The operations list abov
e introduce polymer into the anode buer jar.
When you replace the 1X run buer
, do not add fresh buer to old buer. Discard old
buer, then add fresh buer.
IMPORTANT! Using old 1X buer can lead to loss of resolution and data quality
.
1.
Remov
e the anode buer jar by pulling it down, then twisting it slowly.
2.
Empty the anode buer jar into an aqueous
waste container.
3.
Rinse the anode buer jar using deionized
water.
4.
Rinse the anode buer jar using 1X run
buer.
a.
Add 5 mL 1X run buer to the anode
buer jar.
b.
Tilt the anode buer jar 90°.
c.
Rotate the jar to rinse the interior with buer.
1
Overflow hole
Chapter 2 Prepare the instrument
Fill the anode buffer jar
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
33