3.
Place the Water and Waste reservoirs into the instrument. The reservoirs must be
in the following order from left to right:
1
Buffer reservoir
2
Water reservoir
3
Waste reservoir
4.
Close the instrument door
.
Fill the anode buffer jar
Replace the anode buer at the following times.
Before each group of scheduled runs, or at least every 48 hours
After you replenish polymer
After you replace the capillary array
After you run the Bubble Remove wizard
Every time that you change the buer reservoir
IMPORTANT! The operations list abov
e introduce polymer into the anode buer jar.
When you replace the 1X run buer
, do not add fresh buer to old buer. Discard old
buer, then add fresh buer.
IMPORTANT! Using old 1X buer can lead to loss of resolution and data quality
.
1.
Remov
e the anode buer jar by pulling it down, then twisting it slowly.
2.
Empty the anode buer jar into an aqueous
waste container.
3.
Rinse the anode buer jar using deionized
water.
4.
Rinse the anode buer jar using 1X run
buer.
a.
Add 5 mL 1X run buer to the anode
buer jar.
b.
Tilt the anode buer jar 90°.
c.
Rotate the jar to rinse the interior with buer.
1
Overflow hole
Chapter 2 Prepare the instrument
Fill the anode buffer jar
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
33

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals