Prepare and load sample plates
Prepare sample plates as needed for your application. Refer to the following topics to
help you determine sample placement in wells.
“Default injection positions—48-capillary arrays“ on page 37
“Default injection positions—96-capillary arrays“ on page 39
“Default injection order on plates“ on page 59
1.
Briey centrifuge the plate.
2.
Ensure that each sample is positioned correctly in the boom of its well.
IMPORTANT! If the contents of any w
ell contain bubbles or are not located at the
boom of the well, briey centrifuge the plate, remove the plate from the
centrifuge, and verify that each sample is positioned correctly in the boom of its
well.
3.
Store the plate on ice and protected from light until y
ou prepare the plate
assembly and load the plate in the instrument.
After you load samples in plates, see “Seal and assemble plates“ on page 41.
Default injection positions—48-capillary arrays
2 injections are required to sample all wells on the plate.
Injection 1
Injection 2
96-Well plate,
48 c
apillaries
Chapter 2 Pr
epare the instrument
Prepare and load sample plates
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
37

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals