Perform spectral calibration for sequencing and fragment analysis
A spectral calibration creates a matrix that is used during a run to reduce raw data
from the instrument to the 4- or 5-dy
e data stored in the sample les.
Note: The run times listed below do not include the time for the oven to preheat.
Application Standard Polymer
Capillary array
l
ength
Run module Run Time (min)
[1]
Sequencing Sequencing
standard
POP-7
36 cm Spect36_SeqStd_POP7 60
50 cm Spect50_SeqStd_POP7 115
POP-6
36 cm Spect36_SeqStd_POP6 65
50 cm Spect50_SeqStd_POP6 145
Fragment
analysis
Matrix
standard
POP-7
36 cm Spect36_MtxStd_POP7 50
50 cm Spect50_MtxStd_POP7 60
[1]
The software may take up to 30 minutes to calculate the matrices after the run is complete.
For each dye set–capillary array type–capillary array length combination
After the laser or CCD camera has been realigned/replaced by a service engineer
After the capillary array is changed or replaced
If you see a decrease in spectral separation (pull-up and/or pull-down peaks)
If you alter any condition (dye set, array type, array length, or polymer type)
Spectral
c
alibration
definition and run
time
When to perform
the calibration
Chapter 8 Calibr
ate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
97

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals