Computer and software requirements
Computer/software
Requirement
Computer
IMPORTANT! Do not modify the ins
trument hardware or software without notifying
Thermo Fisher Scientific. Any modifications must be made by Thermo Fisher Scientific
under change control.
The computer provided with the instrument contains validated software and settings.
Do not change the s
tatic IP settings.
For minimum computer requirements, see “Instrument specifications“ on page 142.
IMPORTANT! Do not r
ename the computer after the 3730xl Data Collection Software
5 is installed. The instrument computer has been assigned a unique name. Changing
the name may cause the software to malfunction.
Because of the Ethernet cable connection between the instrument and the computer,
the c
omputer must be located within 9 feet of the instrument.
IMPORTANT! The ins
trument communicates with the computer by Ethernet
connection. Do not make any changes to the computer ethernet/internet connections
during a run or during calibration.
Operating system Operating system: Windows
10 Ent
erprise 2016 LTSB, 64-bit (requires 5 partitions)
Consult with a Thermo Fisher Scientific representative before updating the Windows
operating system or firewall settings.
Instrument firmware Instrument firmware is to be updated only by a Thermo Fisher Scientific representative.
Antivirus software
requirements
No antivirus software is provided with the Data Collection software. The following
applications are compatible with the software:
Symantec Endpoint Protection 12
McAfee Endpoint Security version 10.5
Windows Defender Antivirus (comes as part of the Windows
10 installation)
Other software
CAUTION! Do not ins
tall additional software on the computer other than
antivirus software. Changes to the configured software could void the
instrument warranty and cause the instrument software to be non-operational.
Theory of operation
When DNA samples are prepared for sequencing and fragment analysis on the
instrument, uorescent dy
es are aached to the DNA.
Preparing
samples
Chapter 1 Ins
trument and software description
Computer and software requirements
1
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
17

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals