Check for bubbles in the pump system
1.
Check for bubbles at any point where the pump channels join.
1
2
3
1
Array ferrule tip
2
All tubing
3
Elbow (internal channel) of the lower pump block
2.
If bubbles are present, run the Bubble Remove W
izard. For more information,
see “Typical conditions for using maintenance wizards“ on page 117.
IMPORTANT! When this wizard runs, it introduces polymer into the anode buer jar
.
Replace the anode jar buer after you run this wizard. See “Fill the anode buer
jar“ on page 33.
Chapter 2 Prepare the instrument
Check for bubbles in the pump system
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
35