Clean the polymer delivery pump (PDP) and lower polymer block
Do not expose the polymer blocks to organic solvents.
Do not use sharp or pointed instruments to remove dried polymer from the
polymer blocks.
Do not use water >40°C to clean the polymer blocks.
Clean the exterior 1 time per week, when you replenish polymer.
Flush the polymer delivery pump water trap 1 time per week.
Clean the polymer delivery pump chamber, channels, and tubing 1 time per
month.
Parts of the
polymer delivery
pump
Guidelines
Frequency
Chapter 9 Maintenance
Clean the polymer delivery pump (PDP) and lower polymer block
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide