Consumables and usage limits
Consumable
Storage On-instrument limit
Polymer 2–8°C 7 days
1X run buffer (prepared from
3730 Running Buffer (10X) )
Used in the anode buffer jar and
the buffer reservoir
Shipped at room
temperature. Stable for
1 week.
2–8°C for up to 1 month
48 hours
IMPORTANT! Do not add buff
er to the anode
buffer jar or the buffer reservoir. Always replace
with fresh buffer.
Capillary array Room temperature 300 injections
Important notice regarding use of consumables that exceed
support
ed limits
Life Technologies does not recommend the use of consumables that exceed supported
limits. The recommended limits are designed to promote the production of high
quality data and minimize instrument downtime. Reagent and consumable lifetime
minimum performance are based on testing and studies that use reagents and
consumables that have not exceeded supported limits.
The use of consumables beyond the supported limits may impact data quality or
cause damage to the instrument or capillary array. The cost of repairing such damage
is NOT covered by any Life Technologies product warranty or service plan. Customer
use of expired consumables is at customer's own risk and without recourse to Life
Technologies. For example, product warranties do not apply to defects resulting from
or repairs required due to misuse, neglect, or accident including, without limitation,
operation outside of the environmental or use specications or not in conformance
with Life Technologies instructions for the instrument system, software, or
accessories.
Please see your specic service contract or limited product warranty for exact
language regarding coverage and ask your Life Technologies representative if you
have further questions.
Chapter 1 Ins
trument and software description
Consumables and usage limits
1
16
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals