For more information, see the following sections.
•
“Default injection positions—48-capillary arrays“ on page 37
• “Default injection positions—96-capillary arrays“ on page 39
Run the instrument in auto mode
1.
Select
GA Instruments4 ga37304Instrument name4 Run Scheduler.
2.
At the top of the screen, select Instrument4Instrument name, then select Auto.
Not
e: The Instrument menu is available only when the Run Scheduler is
displayed.
3.
Load the plates into the input stack in the stacker.
See “Move a plate in the Input stack during a run“ on page 64.
The plates will be run in the order in which they are placed in the stacker. The
boom plate runs rst.
4.
Click
(Run).
As the plates are mov
ed to the autosampler, plate barcodes are scanned and their
plate records are used for the run.
IMPORTANT! Ensure that barcodes hav
e been scanned or manually entered for
each plate. If the plate barcode scanned by the instrument does not match a
barcode in the plate record, the plate will be put in the Output Stack.
During an auto run:
•
No plates are listed in the Input Stack
• The currently running plate is listed in the Auto Sampler eld
• Completed plates are listed in the Output Stack
Run the instrument in manual mode
1.
Select
GA Instruments4 ga37304Instrument name4 Run Scheduler.
2.
At the top of the screen, select
Instrument4Instrument name Run Mode,
then select Manual.
Not
e: The Instrument menu is available
only when the Run Scheduler is displayed.
Chapter 5 Run the instrument
Run the instrument in auto mode
5
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
61