5.
Detach the tip of the capillary array from
the pump block, then remove the capillary
array knob and ferrule.
6.
Clean the front surface of the detection cell.
a.
Add one drop of methanol to a sterile
swab or lint-free wipe.
b.
Gently wipe the detection cell.
1
1
Detection cell
7.
Reinstall the tip of the capillary array
, array knob, and ferrule in the pump block.
8.
Push the pump and lower polymer blocks back against the pump panel. Make
sure that the buer valve lever engages the buer pin valve.
9.
Carefully place the detection cell into the detection block. Secure it by rotating
the cam knob 1/4-turn counterclockwise. The pointer on the knob is pointed
down.
10.
Close the detection cell door, oven door, and instrument door.
Perform a short-term shutdown
Perform the short-term shutdown procedure if the instrument will be idle for <1week.
1X run buer
Deionized water
Lab wipes
Gloves
3
2
1
1
Tip of the capillary array
2
Capillary array knob
3
Ferrule
Materials required
Chapter 9 Maint
enance
Perform a short-term shutdown
9
126
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals