Run modules
Fragment analysis run times
Polymer
Capillary Array Length
(cm)
Run Module
Approximate Run Time
(min)
POP-7
36 GeneMapper36_POP7_1 35
50 GeneMapper50_POP7_1 45
36 GS1200LIZ_36_POP7_v2_1 130
50 GS1200LIZ_50_POP7_v2_1 50
Sequence analysis run times
Polymer
Capillary
Array
Length (cm)
Sequencing Run Sequencing Run Modules
Approximate
Run Times
[1]
(minutes)
KB
Basecaller
QV20 CRL
(Bases)
[2]
POP-7
50 Extra long read XLRSeq50_POP7 180 ≥900
BDx_XLRSeq50_POP7
[3]
50 Long read LongSeq50_POP7 120 ≥850
BDx_LongSeq50_POP7
[3]
50 Fast read FastSeq50_POP7 60 ≥700
BDx_FastSeq50_POP7
[3]
36 Standard read StdSeq36_POP7 60 ≥700
BDx_StdSeq36_POP7
[3]
36 Rapid read RapidSeq36_POP7 35 ≥550
BDx_RapidSeq36_POP7
36 Short read TargetSeq36_ POP7 20
[4]
≥400
[4]
C