• Use 3 mm Thermo Fisher Scientic heat seal lm (Cat. No. 4337570). This lm is
3 mm thick before heating and
1 mm thick after heating.
• Do not use heat seal lm that is thicker than 1 mm after heating.
• Do not use heat-seal lm that contains adhesives or metals. These seals can
damage the internal needles that pierce the lm before the injection.
Place plate assemblies into the instrument
1.
Open the stacker drawer.
2.
Open the door of the In Stack tower.
3.
Place the plate assemblies into the stacker
. Ensure that the notched corner of the
plate assembly is at the rear right corner of the stacker.
IMPORTANT! Do not place more than 16 plates in the stacker.
1
Notched corner
Important heat
seal
recommendations
Chapter 2 Prepare the instrument
Place plate assemblies into the instrument
2
44