Dye sets G5 and G5-RCT for fragment analysis
In fragment analysis applications using a 96-capillary array, a small peak in one color
channel that occurs at exactly the same size as a major peak in another color channel
can be incorrectly identied as an allele. This phenomena is referred to as "cross-talk"
between the dierent spectral channels.
The G5-RCT (Reduced Cross Talk) dye set has been optimized to minimize cross talk
peaks.
RCT dye sets reduce potential crosstalk peaks, but also cause a decrease in overall
signal intensity.
RCT dye sets slightly reduce signal-to-noise ratio.
Because of the decrease in overall signal intensity, a higher dynamic range is
available when using RCT dye sets.
The spectral calibration for an RCT dye set is specic to the capillary array on
which is it run. Do not ovveride spectral calibration when you use an RCT dye
set.
Fragment analysis with a 96-capillary array
Applications with a high dynamic range (where the signal intensity of large
peaks is much higher than the signal intensity of small peaks)
Fragment analysis
using a 96-
capillary array:
cross-talk peaks
Reduced cross-
talk—Dyeset G5-
RCT
When to use RCT
dye sets
Appendix D Dye sets: G5, G5-RCT, Any4Dye, Any4Dye-HDR, and Any5Dye
Dye sets G5 and G5-RCT for fragment analysis
D
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
149

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals