Dye set G5 created from matrix standard set DS-33
Best practice is to perform a spectral calibration every time you install a new
capillary array
.
There may be circumstances where you want to reuse a calibration that was
generated using a dierent capillary array.
Note: You must select the calibration before data is collected. After data is
collected, you cannot reapply a dierent spectral calibration.
Do not reuse a previous calibration when you use the G5-RCT dye set. Because of
the way the spectral calibration is generated, it is specic to the capillary array
used to generate the calibration.
1.
In the navigation pane of the Data Collection Software, select
GA
Instruments4 ga37304 Instrument name4 Spectral Viewer.
Failed calibrations are marked with an asterisk *.
2.
Select the dy
e set of interest.
3.
Click the List of Calibrations list in the lower right-pane.
Example of a
passing fragment
analysis spectral
calibration
Select a previous
c
alibration as the
active spectral
calibration
Chapter 8 Calibr
ate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
111

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals