Applied Biosystems 3730xl User Manual

Applied Biosystems 3730xl
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
8.
Fill, then assemble the reservoirs.
Buffer reservoir assembly Water and waste reservoir assemblies
1. Add 80 mL 1X run buffer to the
Buffer reservoir.
2. Assemble the reservoir assembly as
shown below.
1
2
3
4
5
6
1
Retainer
2
Septum
3
Reservoir cap
4
Reservoir
5
Heated plate base
6
Power cable
1. Add 80 mL high-quality deionized
water to each reservoir.
2. Assemble each reservoir assembly
as shown below.
1
2
3
4
5
1
Retainer
2
Septum
3
Reservoir cap
4
Reservoir
5
Plate base
9.
To prevent damage to the capillary array, inspect each reservoir for the following
conditions.
Septa t snugly and ush on the reservoir cap
Rubber gasket surrounding the reservoir cap is seated correctly
1
1
Rubber gasket not seated correctly
Chapter 2 Prepare the instrument
Prepare buffer and fill the reservoirs
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
31

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals