8.
In the Instrument Protocol eld, click the eld, select , then select a protocol.
For information on instrument protocols, see Chapter 7, “Create and manage
instrument protocols, run modules, and analysis protocols“
.
9.
For the remaining elds, select click the eld, select
, then select an
appropriate item for y
our application.
The seings in these lists are provided by the secondary analysis software. For
information on these seings, see the software help in the secondary analysis
software.
10.
Fill in the remaining rows by manually entering and selecting information, by
using Fill Down commands, or by copy/pasting.
Fill Down Copy/paste
Select the entire row, select Edit4Fill
Do
wn Special, then select a fill down
option.
For 48-capillary instruments, select Fill
down Special (48 Cap).
For 96-capillary instruments:
• 96-well plate: Select Fill Down.
• 384-well plate: Fill down Special
(96 Cap).
• If you copy multiple cells, select the
same number of corresponding
target cells before you paste.
• You can copy/paste within one
protocol.
Note: Use the Duplicate function in
the Plate Manager to copy
information from an existing
protocol to a new protocol.
11.
Click OK.
Chapter 4 Create results groups and plate records using default settings
Create a plate record
4
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
57