The higher the SatSeconds limit is set, the longer it takes for the upper or lower limit alarm to occur. Set the
SatSeconds limit appropriately by taking into consideration the patient’s condition. For example, select a
higher setting for monitoring an active patient whose %SpO2 values tend to fluctuate greatly.
The upper or lower limit alarm will occur even when the whole indicator has not changed color if a detected
%SpO2 value deviates from the acceptable range three or more times within 60 seconds.