Troubleshooting
64
Alarm name
Pulse search
Sensor discon-
nect
Sensor off
Interference
Defective
sensor
Internal error
Cat-
egory
Pulse
Condition causing
the alarm
This alarm will occur
when it is difficult to
detect a pulse.
This alarm will occur
if the sensor is not con-
nected (or the connec-
tor has come off).
This alarm will occur
if the sensor attached
to the patient has
come off.
This alarm will occur
due to strong interfer-
ence of noise, etc.
This alarm will occur
if the sensor has be-
come defective and
cannot be recognized.
This alarm will occur
if the unit has encoun-
tered a problem.
The unit during the
alarm condition
“0” will be dis-
played.
“– – –” will be dis-
played.
“– – –” will be dis-
played.
“– – –” will be dis-
played.
“0” will be dis-
played.
“0” will be dis-
played.
Resetting the
alarm
The alarm will be
reset automatically
when pulses are
detected.
The alarm will be
reset automatically
when the sensor is
connected prop-
erly.
The alarm will be
reset automatically
when the sensor is
attached properly.
The alarm will be
reset automatically
when the noise has
decreased.
Replace the defec-
tive sensor with a
functional one. (If
the connector is re-
moved, “Defective
sensor” alarm will
be replaced with
“Sensor discon-
nect” alarm.
Turn the pulse
function off.
Alarm
silence
2 min.
Reset
2 min.
2 min.
2 min.
Switched
to warning
informa-
tion to the
user.
Message
Pulse Oximeter
No pulse detected.
Pulse Oximeter
Check the sensor
connection.
Pulse Oximeter
Confirm proper
placement of the
sensor.
Pulse Oximeter
No pulse detected.
Pulse Oximeter
No pulse detected.
Pulse Oximeter
No pulse detected.
Priority
High
Medium
Medium
Medium
Medium
Low