Troubleshooting
MAINTENANCE
65
Alarm name
Sensor module
position alarm
Filter cover
alarm
Fan alarm
System failure
alarm
Cat-
egory
Others
Condition causing
the alarm
This alarm will occur
if the sensor module is
left removed for 15
seconds or longer af-
ter calibration, or if it
remains in a position
other than the calibra-
tion position or the
normal position for 15
seconds or longer.
This alarm will occur
if the filter cover is left
open.
This alarm will occur
if the fan isn’t attached
properly.
This alarm will occur
if the fan cover or the
middle board isn’t at-
tached properly.
This alarm will occur
if the fan rotates at a
lower rate than speci-
fied.
This alarm will occur
if a wire of the incuba-
tor air temperature
sensor (for control)
has snapped or short-
circuited.
This alarm will occur
if a wire of the incuba-
tor air temperature
sensor (for display)
has snapped or short-
circuited.
This alarm will occur
if a wire of the incuba-
tor heater has snapped
or short-circuited.
This alarm will occur
if a wire of the radiant
warmer heater has
snapped or short-cir-
cuited.
The unit during the
alarm condition
The fan will con-
tinue normal op-
eration.
The fan will be
stopped.
The incubator
heater will be
turned off.
The fan will con-
tinue normal op-
eration.
The incubator
heater will be
turned off.
The fan will be
stopped.
The incubator
heater will be
turned off.
The fan will ro-
tate at an avail-
able rate.
The incubator
heater will be
turned off.
The incubator
heater will be
turned off.
Incubator air
temperature con-
trol will be con-
tinued.
The incubator
heater will be
turned off.
The radiant
warmer heater
will be turned off.
Resetting the
alarm
Replace the sensor
module to its
proper position.
The alarm will be
reset automatically
when the filter
cover is closed.
Attach the fan
properly, and then
turn the power on
again.
The alarm will be
reset automatically
when the fan cover
or the middle
board is attached
properly.
Impossible to reset
the alarm.
Impossible to reset
the alarm.
Alarm
silence
2 min.
2 min.
15
min.
2 min.
10
min.
10
min.
10
min.
10
min.
10
min.
Message
Place the sensor
module back in its
proper place.
Close the filter
cover.
Disconnect power
and check the fan
is properly at-
tached.
Check the fan
cover is properly
attached.
Revolving speed
of the fan is de-
creased. (E4)
Abnormality in
the incubator air
temperature sen-
sor is detected.
(E262)
Abnormality in
the incubator air
temperature sen-
sor is detected.
(E263)
Incubator heater
abnormal. (E7)
Warmer heater
abnormal. (E9)
Priority
Medium
Medium
Low
Medium
Medium
Medium
Medium
Medium
Medium