Troubleshooting
66
Alarm name
Power failure /
internal error
alarm
Cat-
egory
Others
Condition causing
the alarm
This alarm will occur
if the power supply is
interrupted due to
power failure, a dis-
connected power plug,
a broken power cord,
a tripped breaker or
some other cause, or if
some internal error (a
defective CPU or a de-
fective control board)
has occurred.
The unit during the
alarm condition
No control
All the heaters
will be turned off.
Resetting the
alarm
Impossible to reset
the alarm
Alarm
silence
Impos-
sible
to
silence
the
alarm
Message
No message (The
power failure
alarm indicator
will come on.)
Priority
High

Table of Contents