Troubleshooting
72
(1) When any of the following messages is displayed on the screen:
q “Abnormality in yellow skin temperature probe”
YES
NO
The alarm is still displayed even
after the skin temperature probe is
replaced with a functional one.
Failure of the detection board
Failure of the skin temperature
w “Check the connection of the yellow skin temperature probe.”
YES
NO
The yellow skin temperature probe is
connected securely to the skin
temperature probe connecting port 1.
Failure of the detection board
Connect the yellow skin temperature
probe securely to the skin
temperature probe connecting port 1.