Troubleshooting
MAINTENANCE
73
e “Check that the humidity chamber is attached properly.”
NO
NO
NO
YES
NO
YES
YES
YES
The humidity chamber is fit in.
The boiler cap is attached to the
humidity chamber.
Fit in the humidity chamber.
Attach the boiler cap to the humidity
chamber.
The humidity chamber is inserted
completely into the unit.
Insert the humidity chamber
completely into the unit.
The connector of the position
detection board is connected
properly.
Connect the connector of the
position detection board securely.
Failure of the position detection board
Failure of the control board