54733-10BA2L
3 – 304 02.2017
1 HART
®
Communication Volume 3
1.5.1 HART
®
Communicator
The HART
®
Communicator Model 375/475 allows you to archive
parameters sets. For this purpose, the parameter set first has to be
saved using the software button SAVE from the detector to the SD
card of the HART
®
Communicator. The stored data can then be
archived to a PC via SD card or via infrared interface. First, prepa-
ratory actions have to be performed in the Online menu. The follow-
ing sequence shows how the parameter set can be transferred from
the detector to a PC via SD card.
1. Enter the file name of the parameter set for archiving in the Tag
parameter.
Command: Tag
Execution:
Online via the HART
®
Communicator in the Identification
menu, select the Tag parameter and enter a name. Up to 8
characters.
2. Load current measurement parameters to the calibration
parameters.
Command: Recall
Execution:
Online via the HART
®
Communicator in the Cal Parameter
menu, select the command Recall.
3. Save parameter set to the HART
®
Communicator.
Command: SAVE
Execution:
Online via the HART
®
Communicator, push the software button
SAVE on the display.
Info:
The parameter set is now also available offline on the HART
®
Communicator.