SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 305
Volume 3 1 HART
®
Communication
3
4. Transfer parameter set to PC.
You need the program "Easy Upgrade Utility" for this step. This
program must be installed earlier on the PC.
Execution:
– Remove SD card from 475 and insert it into the SD card
reader of your PC.
– Start Easy Upgrade Utility.
The following steps describe the sequence in "Easy Upgrade
Utility":
– ("Update PC" recommended)
– Connection type: Card Reader
– "Connect"
– Select "More options ...".
– Select "HART configuration" tab.
– Select a file in the right box and upload it to the PC data-
base in the left box.
The parameter set can now be displayed, or printed, by double-
clicking.
For more information about "Easy Upgrade Utility" please refer
to the Online help.
To transfer already archived files to the detector:
1. Transfer the file from PC to SD card using "Easy Upgrade Util-
ity".
2. Insert SD card again in the HART
®
Communicator.
3. Select the parameter set in the offline mode of the HART
®
Communicator and transfer it to the detector with the Send
command.
The detector is now ready for measurement.
To update the data, you have to restart the HART
®
Communicator.
For archiving with other hosts please see the following chapter.