1 HART
®
Communication Volume 3
1.5.2 PDM (Siemens Simatic PDM)
The PDM can also store data in the offline menu. Again, you have to
select Recall on the Cal Parameter menu.
When restoring the data to the detector, it is necessary to transfer
the data twice to the detector. This ensures that the data will be
properly enabled in the detector.
1.5.3 AMS (DeltaV Emerson Process)
The AMS can save data in the offline menu only from version 10 and
above. Again, you have to select Recall on the Cal Parameter
menu.
However, safe transfer of data from the offline menu to the detector
can only be guaranteed by transmitting each parameter individu-
ally. You have to check whether the corresponding value has actu-
ally been transmitted properly. Then in addition you have to select
Restore in the Cal Parameter menu to transfer the calibration val-
ues into the measurement parameters.