EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB 480 - Page 345

BERTHOLD TECHNOLOGIES LB 480
450 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 345
Volume 3 Menu Structure
3
•MANUAL GDA (Gas Density Adjustment)
If the container in operation is under gas pressure (> about 2
bar) and the calibration cannot be carried out under operating
pressure, then it is possible to calculate in advance the count
rates that should be obtained under operating pressure. Addi-
tional parameters are displayed which are required for the com-
pensation.
Gas measuring path
Setup Gas Density
Gas Density at Op
Please note that with a multi-point calibration the GDA can only
be used when the same gas density is available at all calibration
points during calibration. Only the calibration point at 100% is
excluded, provided that the measurement path in the container
is completely filled with product.
•AUTO GPC (Gas Property Compensation)
IMPORTANT
This function requires special knowledge and should be enabled
only by a Berthold service engineer, or by a specially trained per-
son.
If the vessel is in operation under gas pressure (> about 2 bar)
and the gas pressure is not constant, then a continuous gas
density compensation is advisable. An additional measurement
is required which is arranged above the liquid level measure-
ment and constantly measures the gas density in the container.
Additional parameters required for this calibration appear on
the Calibration menu.
A further prerequisite for GPC is that the source used for level
measurement as well as for compensation measurement con-
tain the same isotope. This means, if a level measurement is
carried out using Cs-137, then the compensation measurement
must also be carried out with Cs-137.
GPC cannot be used for a level measurement with point source
and point detector.
4 ... 20 mA
HART
Detector for GPC
compensation
measurement
RS 485
Detector for level
measurement

Table of Contents

Related product manuals