Menu Structure Volume 3
4 Gas Measuring Path Visible only if Compensation Mode is set to MANUAL GDA.
Enter the average measurement path through the gas-filled
room.
With point sources and rod detectors:
The path from the point source to the center of the measuring
range. However, only the path through the container, i.e. in the
gas-filled room.
With point detectors and rod sources:
The path from the point source to the center of the measuring
range. However, only the path through the container, i.e. in the
gas-filled room.
In oblique irradiations with point source and point detector:
The path from the source to the detector. However, only the path
inside the gas-filled room.