BERTHOLD TECHNOLOGIES GmbH & Co. KG
Code
Error message
Error Handling
N=Normal
S=Sensitive
Error Log
HART
®
Digital Out
Fault current
SHUTDOWN
Slave Status
Master Status
Self-repairing
101 HW module miss-
ing or not tested
N xxxxxbb–
S xxxxbb–x
102 Device data-set
error
N xxxxxbb–
S xxxxxbb–
103RAM Error N xxxxxbb–
S xxxxxbb–
104Device Error N xxxxxbb–
S xxxxxbb–
105 Real time clock
not valid
N xxx uu–
S xxxx bb–
106Test mode active N gu–
Sgu–
107 Watchdog reset N x g g x
S xxxxxbb–
108 Safety parameter
invalid
N gg–
S xxxx bbx
200Data flow N xxxxxbb–
S xxxxxbb–
201 Error by analog
input calibration
N xxx uu–
S xxxx uux
202 Clock signal devi-
ation
N xxxxxbb–
S xxxxxbb–
300Data flow N xxxxxbb–
S xxxxxbb–
301 Error by ADC cal-
ibration
N xxx uu–
S xxxx uu–
302 Error by DAC cal-
ibration
N xxx uu–
S xxxx uu–
303Supply 5.0V N xxxx bbx
S xxxx bbx
304Reference 2.0V N xxxx bbx
S xxxx bbx
305Reference 2.5V N xxxx bbx
S xxxx bbx
306 ERROR GND CPU
ST9
N xxxx bbx
S xxxx bbx
307 No impulses in
measuring chan-
nel
N xxxx bbx
S xxxx bbx
308 No impulses in
control channel
N xxx uux
S xxxx bbx
309 No impulses in
auxiliary channel
N xxx uux
S xxxx bbx
310 Impulse differ-
ence measuring
channel
N xxxx bbx
S xxxx bbx
311 Impulse differ-
ence control
channel
N xxxx bbx
S xxxx bbx
312 Impulse differ-
ence auxiliary
channel
N xxxx bbx
S xxxx bbx
313 Instable pulse
rate
N ggx
S ggx
314 Threshold of
measurement
channel 1
N xxx uux
S xxxx bbx
315 Threshold of
measurement
channel 2
N xxx uux
S xxxx bbx
316 Threshold of con-
trol channel 1
N xxx uux
S xxxx bbx
317 Threshold of con-
trol channel 2
N xxx uux
S xxxx bbx
318 Threshold of aux-
iliary channel 1
N xxx uux
S xxxx bbx
319 Threshold of aux-
iliary channel 2
N xxx uux
S xxxx bbx
320 HV voltage N x x x u u x
S xxxxxbb–
321 Generated HV
voltage
N xxxx bbx
S xxxxxbb–
322 HV reached its
limit value
N xxxx bbx
S xxxx bbx
323 HV average is
20% lower than
default HV
N xxxx bbx
S xxxx bbx
324 HV average is
40% higher than
default HV
N xxx ggx
S xxxx uux
325 Lower PMT cur-
rent limit is
exceeded
N x uux
S x uux
326 Upper PMT cur-
rent limit is
exceeded
N xxxxxbb–
S xxxxxbb–
327 Temperature
sensor deviation
N xxx uux
S xxxx bbx
Code
Error message
Error Handling
N=Normal
S=Sensitive
Error Log
HART
®
Digital Out
Fault current
SHUTDOWN
Slave Status
Master Status
Self-repairing