8 Error Handling Volume 3
613 Inherited Message A new software applica-
tion has been installed
and a previously stored
fault is not existent in
the new application.
Clear the fault memory with Reset
Error Log.
702 CLoop Malfunction
Loose connection in
the current loop.
Impedance in the cur-
rent loop
>500Ω
Fault in the power out-
put of the LB 480
If a loose connection or a too high
impedance can be ruled out, the probe
must be returned to the factory for
repair.
703 CLoop Monitoring Disabled The monitoring of the
current output is turned
off. Message always
appears when restarting.
Enable the monitoring on the I/O Setup
menu, unless there are reasons to keep
the monitoring switched off.
901 Signal Unlocked The detector has been
unlocked with the pass-
word.
The warning message was generated as
a result of the setting in Signal
Unlocked, see chapter 2.16, page
3-329.