SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 435
Volume 3 8 Error Handling
3
407 Remote device error A slave returns an error
message.
Go to the Diagnostics menu and check
in the local slave table which error mes-
sage is indicated for which slave.
504 Inherited Message A new software applica-
tion has been installed
and a previously stored
fault is not existent in
the new application.
Clear the fault memory with Reset
Error Log.
601 License Key Error The license key is invalid
or does not match the
application.
Check and correct license key. If neces-
sary, consult BERTHOLD
TECHNOLOGIES.
603 Measuring Error check <Error Status> During decay compensa-
tion, an error was
detected in the measure-
ment parameters.
For more information on the cause of
the error please go to the Diagnostics
menu under Error Status.
604 Decay Compensation Error The decay compensation
could not be performed.
The error can be eliminated only by
replacing the detector electronics.
605 Source Exchange Based on the criteria
entered, it was found
that the source is too
weak.
Replace source at the next opportunity.
Please contact the manufacturer.
606 Radiation Interference Based on the criteria
entered radiation inter-
ference was detected.
Check whether welding seam tests were
performed, or whether the criteria are
set incorrectly or not optimally.
607 RID Interference Message in this software
is not active.
608 Inherited Message A new software applica-
tion has been installed
and a previously stored
fault is not existent in
the new application.
Clear the fault memory with Reset
Error Log.
609 Rate Factor exceeds its limit The automatic gas den-
sity compensation is
enabled. During the
measurement, the den-
sity compensation factor
has exceeded the per-
missible range.
Check the calibration and the settings of
the gas density compensation.
610 Inherited Message A new software applica-
tion has been installed
and a previously stored
fault is not existent in
the new application.
Clear the fault memory with Reset
Error Log.
611 Inherited Message A new software applica-
tion has been installed
and a previously stored
fault is not existent in
the new application.
Clear the fault memory with Reset
Error Log.
612 Inherited Message A new software applica-
tion has been installed
and a previously stored
fault is not existent in
the new application.
Clear the fault memory with Reset
Error Log.
Code
Error message Error reason Error Handling