EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB 480 - Page 443

BERTHOLD TECHNOLOGIES LB 480
450 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 443
Volume 3 9 Setup Protocol
3
Path Parameters Unit Standard SETUP
Device Config Setup Signal Condition Signal Dependency
Response Mode DISABLED
RAPID SWITCH
RAD INTERFERENCE
DISABLED
Io Factor 1.5
Waiting Time s 60
RI Sigma 10
Meas Delay Time s 4
Device Config Setup I/O Setup Current Output
Current Loop Monitoring DISABLED
ENABLED
ENABLED
Loop Alarm Type HIGH
LOW
HOLD LAST VALUE
VALUE
HIGH
Error Current Value mA 22.00
Current Lower Limit mA 3.80
Current Upper Limit mA 20.50

Table of Contents

Related product manuals