Manual Operation
3-53
Figure 3-79
When conducting the APG current measurement calibration, each
calibration point has to enter at least 4 Arabic numerals to ensure the
power supply accuracy after calibration.
Calibration Procedure (Model 62180D-600):
1. In Calibration page, tap ā€œHost Vā€ to select APG I MEAS and list the APG current
calibration items as shown in Figure 3-80.
Figure 3-80