Responding to Error Messages
102
Chapter 4 FAQ
If the error occurred during milling
Press the operation button on the built-in panel.
Milling will resume.
If the error occurred while the machine was on standby
Hold down the operation button on the built-in panel.
This will clear the error.
Situation/Error Cause 2
There is a possibility that the ATC magazine is out of position.
Procedure
Perform automatic correction.
" P.65 "Correcting the Milling Machine"
If the Error Occurs Again
The collet may have worn out. Replace the collet. If the error occurs again even after you replace the collet, replace
the spindle unit.
" P.62 "Situations Requiring Part Replacement"
"1023-0000 to 1028-0000" Milling data error.
• 1023-0000: Milling data error. The number of parameters is incorrect.
• 1024-0000: Milling data error. The parameter is out of range.
• 1025-0000: Milling data error. A wrong command is detected.
• 1026-0000: Milling data error. The address is not dened.
• 1027-0000: Milling data error. The parameter is not dened.
• 1028-0000: Milling data error. The operation cannot be executed.
Situation/Error Cause
There may be a problem with the milling data received from the computer.
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Review the CAM settings and the shape specied in the CAD data.
If there are no problems with the milling data
Restart the computer, and then perform milling again.