Responding to Error Messages
106
Chapter 4 FAQ
"103B-0000" The automatic correction is not yet nished.
Situation/Error Cause
• Automatic correction has not been performed.
• Automatic correction was cancelled before it could nish, and the machine was left in this state.
• Automatic correction was not performed after updating the rmware to a version that required automatic
correction to be performed again.
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Perform automatic correction.
" P.65 "Correcting the Milling Machine"
Situation/Error Cause 2
The VPanel and machine rmware versions may not match.
Procedure
Download the latest versions of VPanel and the machine's rmware, and then install these
versions.
DGSHAPE Corporation website (http://www.dgshape.com/)
"103D-0000" Milling data error. The milling bur cannot reach the milling position.
Situation/Error Cause
The milling bur does not reach the milling position in the Z direction because the milling bur is too short or the
angle of the A axis or B axis is too large.
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Review the milling bur length and the position of the milling bur holder.
" P.45 "Step 3: Setting the Milling Bur"
C
Review the CAM settings and decrease the angles of the A axis and the B axis.