Responding to Error Messages
98
Chapter 4 FAQ
If the Error Occurs Again
The collet may be deformed. In this case, replace the collet.
" P.62 "Situations Requiring Part Replacement"
"101D-0010" The cleaning tool cannot be released.
Situation/Error Cause
The returning of the cleaning tool failed. The inside of the collet or the ATC magazine might be dirty.
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Click [Open collet] in VPanel.
The collet will open.
" P.14 ""Maintenance" Tab"
C
Remove the cleaning tool.
D
Clean the ATC magazine.
" P.59 "Cleaning after Milling Finishes"
Situation/Error Cause 2
The collet and cleaning tool are axed together and cannot be separated. The inside of the collet might be dirty.
Procedure
Retighten the collet.
" P.71 "Retightening the Collet"
If the Error Occurs Again
The collet may be deformed. In this case, replace the collet.
" P.62 "Situations Requiring Part Replacement"