OperationModel 2500 148
© 2015-2017 Distek, Inc. All Rights Reserved
2 The
Select V essel to C a libra te
screen is displayed with vessel position 1 automatically
selected as shown.
Select Vessel to Calibrate screen
3 Place the TempChek probe into vessel 1 and wait for temperature reading to equilibrate.
4 Once the temperature has equilibrated, select
Ca libra te
to start vessel 1 calibration as
shown.
Select Calibrate to calibrate vessel 1
5 Observe that the
V essel Tem p °C
closely matches (±0.05°C) the TempChek current
temperature reading and that the
Lo-C a l
parameter is updated. Repeat steps 4 and 5 if it
does not meet the criteria.
6 Once satisfied with the result, continue the calibration by selecting vessel position 2 and
repeat steps 3 to 5.
7 Repeat the above steps for the remaining vessel positions.