OperationModel 2500 149
© 2015-2017 Distek, Inc. All Rights Reserved
8 Once the Lo-Cal is complete, exit the shaft calibration screen by selecting
Done
.
9 Set the bath temperature to 41°C and wait for the vessel temperature to equilibrate.
10 Repeat steps 1 to 5 to start the Hi-Cal calibration.
11 Observe that the
V essel Tem p °C
closely matches (±0.05°C) the TempChek current
temperature reading and that the
Hi-Ca l
parameter is updated. Repeat steps 4 to 11 if it
does not meet the criteria.
12 Repeat the above steps for the remaining vessel positions.
13 Once the Hi-Cal is complete, exit the shaft calibration screen by selecting
Done
.