Create New Method Screen
You can also create a new method by using the Settings tab, then the Methods
icon.
If you enter the wrong data, a popup message alerts you about the range of a
particular parameter.
A summary of settings for each parameter and the function of each command button is shown
below:
Method Name can be up to
16
characters.
Available apparatus are
U SP 1 Ba sket, U SP 2 Pa d d le, M o dified
Ba sket, M odified Pa dd les, Intrinsic.
Temperature range is from 20.0°C to 65.0°C .
Volume range is from 500mL to 1000mL.
Enable which vessels will be used (available only to 2500 RTD
and Select).
RPM range is from 30 to 300 RPM.
Maximum Duration time for the test is
745
hours and
59
minutes.
Number of Sampling Times is
0
to
32
.
The number of sampling time points in general should be
equal to your test duration time.
Stagger drop time is from
20
to
120
seconds. Enter 0 for
simultaneous sample introduction.
Infinity Duration range from
8
hours and
59
minutes. Leave
setting at 0:00 if it is not required.
Infinity Sampling Time in general should be equal to your
infinity duration time.
Infinity RPM range is from
30
to
300
RPM. Generally the