Sweep Frequency Response Analyzer (SFRA) User Guide
5-14 72A-2570-01 Rev. K 07/2011
Line 3 contains the start tag of the first transformer nameplate and
must also have an end tag (line 54).
Line 55 shows the start of another transformer, which ends on line
106.
If another transformer were added, it would appear after line 106 and
before </transformerNameplates> on line 107.
Merging Settings Files
The method for merging transformer settings files described here applies to all
other settings files.
The following example concerns two SFRA users who want to share XML
files.
Figure 5.12 shows a simplified transformer settings file from User A (as
viewed in Notepad):
Figure 5.12 Transformer Settings File — User A
Use caution when merging, because an incorrect merge could corrupt
your files and cause unpredictable results. Be especially careful to avoid
duplicate entries in any settings file.
The manual merge described here is prone to cut-and-paste errors. Back
up all data and settings files before performing a merge operation.
<?xml version="1.0" encoding="utf-8"?>
<transformerNameplates>
<transformerNameplate version="1">
<manufacturer>ABB
</manufacturer>
<serialNumber>1234
</serialNumber>
</transformerNameplate>
<transformerNameplate version="1">
<manufacturer>GE
</manufacturer>
<serialNumber>678
</serialNumber>
</transformerNameplate>
</transformerNameplates>