Instructions for use Fabius Tiro SW 3.n 73
Assembly and preparation
Table with recommended hose configurations
*
* The resistance of the breathing system and connected accessories must be observed.
Adults Pediatric patients Neonates
Filter or HMEF with connection for sample line between Y-piece and
patient:
Or
Filter at inspiratory port and expiratory port, connection for sample
line at Y-piece:
Filter at inspiratory port, connec-
tion for sample line as close to
patient as possible:
Lateral connections for the sam-
ple line support CO
2 measure-
ment and help to flush the dead-
space between Y-piece and hose
adapter.
LQVS
H[S
LQVS
H[S
21458
WARNING
Risk of underpressure in the lungs
If filters are blocked, the sample gas flow can
immediately cause underpressure in the
lungs.
When ventilating pediatric patients and
neonates, do not use HMEF or other filters at
the Y-piece in connection with a hose adapter
that have a patient-side connection for a
sample line.