Communications
Figure 86 Analog & Digital I/O Individual Pin Descriptions
CAUTION
The analog and digital inputs and outputs are rated to CAT I.
Exceeding 12 VDC or drawing greater than 400 mA on a single output or a total
greater than 2 A across the eight outputs can permanently damage the
instrument and void the warranty.
4.5 Logging Data
When the user receives the instrument from the factory it will have a default set of parameters already
setup in the internal data logger. These select few parameters have been chosen for their relevance in
assisting in troubleshooting the instrument.
4.5.1 Configure Instrument Internal Logging
In order to log data the user must first specify a data logging interval. This is how often data will be
logged to the USB memory stick. The user can log a maximum of 12 parameters. These parameters can
be selected by name or by parameter number using the parameter list as a reference (refer to
Table 33).
4.5.1.1 Data Log Setup Numeric
Procedure
1. Open - Main Menu Communications Menu Data Logging Menu (refer to Section 3.4.23).
2. Select - Data Log Interval - (adjust to the desired value) - Accept.
3. Open - Data Log Setup -Numeric - (select the parameter numbers user wishes to log).