EasyManua.ls Logo

Ecotech Acoem Serinus 30 - General Settings Menu; Figure 50 - General Settings Menu Screen

Ecotech Acoem Serinus 30
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus® 30 User Manual 3.3
Page 64
3.4.8 General Settings Menu
Main Menu General Settings Menu
Figure 50 General Settings Menu Screen
Decimal Places
Select the number of decimal places (0 - 5) used for the concentration
displayed on the home screen.
Conc. Units
Sets the concentration units (ppm, ppb, ppt, mg/m3, μg/m3 or
ng/m3).
Reference Temperature
[Gravimetric Units]
This option only appears if concentration units are set to gravimetric
(mg/m3, μg/m3 or ng/m3). Select either 0 °C, 20 °C or 25 °C. This sets
the standard temperature used from conversion for measured
volumetric values.
Temperature Units
Select the units that temperature will be displayed in (Celsius,
Fahrenheit or Kelvin).
Pressure Units
Select the units that the pressure will be displayed in (torr, PSI, mBar,
ATM or kPa).
Flow Units
Select the units that the sample flow will be displayed in (slpm or
L/min).
Date
Displays the current date and allows users to edit if required.
Time
Displays the current time and allows users to edit if required.
Backlight
Select the length of time the screen and keypad backlight remain ON
after a button press. The setting Always Off means the backlight never
turns ON; the setting Always On means the backlight never turns OFF
and the setting Daytime means the backlight will turns ON from 7 am
to 7 pm or 30 minutes after any operation.
Home Screen
This field allows the user to display concentrations on the Home
Screen in two formats. The first is Inst. only which displays only the
instantaneous concentration reading, the second is Inst & Avg which
displays both instantaneous and average concentration on the Home
Screen. The average is measured over the time period set in
Measurement Settings Menu (refer to Section 3.4.9).
Char 0 has Slash
When enabled, the instrument will display the zero character with a
slash (0) to differentiate it from a capital ‘O’.

Table of Contents

Related product manuals