Serinus® 30 User Manual 3.3
Page 48
Measurement Settings Menu
Background Interval: 24 hrs
Calibration Menu
Span Comp: Disabled
Diagnostics Menu
Press/Temp/Flow Comp: ON
Diagnostic Mode: Operate
Control loop: Enabled
The instrument must be operated and maintained in accordance with this User Manual.
The Serinus 30 analyser is designated U.S. EPA reference method with or without the following
options/items:
Internal pump
Rack mount assembly
Optional Ethernet Port
2.5 EN Type Approval Set-Up
The Serinus 30 has been certified to TÜV performance standards for Continuous Ambient Air Quality
Monitoring Systems. The certificate number is TÜV 936/21221977/D. The Serinus 30 must be used
under the following conditions to meet EN requirements:
Range
0 - 100 ppm
Ambient Temperature
0 - 30 °C
Instrument Settings
The instrument must be operated and maintained in accordance with this user manual.
The following menu selections must be used:
Measurement Settings
Background interval: 24 hrs
Calibration Menu
Span Comp: Disabled