Note: In Figure 80 all the IP addresses are taken as an example. The WAN IP addresses are
normally provided by your ISP. Whereas, the LAN IP addresses can be set manually to any range
which is within the subnet of the Modem/Router/Switch.
Use a cross-over LAN cable to connect the instrument directly to a computer, or a standard LAN cable
for connection to a Modem/Router/Switch as shown in Figure 80. The computer could be connected
to the Modem/Router using either CAT5 cable or a wireless connection, but the instrument must be
connected using CAT5/6 cable.
4.3.1 Reading Network Port Setup
To read the current network port settings perform the following steps:
Procedure
1. Open - Main Menu Communications Menu Network Menu.
2. Select - Start-up Mode Read IP - Accept.
3. Manually use the power switch ON the rear of the instrument to turn the power OFF. Leave the
instrument OFF for 10 seconds before turning the power back ON.
4. Open - Main Menu Communications Menu Network Menu.
5. The current network port settings will now be displayed on the screen.
6. When viewing is complete select - Start-up Mode Normal - Accept.
4.3.2 Setting Network Port Setup
Below is an example of how to setup the network port.
Procedure
1. Open - Main Menu Communications Menu Network Menu.
2. Select - Start-up Mode Set IP - Accept.
3. Edit - IP Address - (Change the IP address to the address the user wishes to use within the
Modem/Router/Switch Subnet).
4. Edit - Netmask - (Change the Netmask to the setup specified by the Modem/Router).
5. Edit - Gateway - (Change the Gateway to the setup specified by the Modem/Router).