4. Edit - (Change one of the storage locations “Parameter 1 - 12” to the parameter number user
wishes to log) - Accept.
4.5.1.2 Data Log – Text
Procedure
1. Open - Main Menu → Communications Menu → Data Logging Menu (refer to Section 3.4.23).
2. Select - Data Log Interval - (adjust to the desired value) - Accept.
3. Open - Data Log Setup –Text - (select the names of the parameters user wishes to log).
4. Select - (Change one of the storage locations “P1 - P12” to the parameter text user wishes to log)
- Accept.
4.6 Using Airodis Software to Download Data
4.6.1 Connecting the Instrument to a PC
The instrument can communicate with a PC using RS-232 (Serial), TCP/IP (Network), Bluetooth or USB.
Serial, Bluetooth and network communications do not require additional drivers. If the user wishes to
connect using a USB cable, the driver must first be installed.
4.6.1.1 Connecting Over USB
If the user wishes to connect using USB, then they will need to first install the Serinus USB driver.
Power ON the instrument and connect it to a PC with a USB cable. The user should receive a prompt if
the driver needs to be installed. If not, open Device Manager (Under “System” in Control Panel), find
the device and select “Update Driver Software”.
Figure 87 – Installing Driver Software (Device Manager)