Serinus® 30 User Manual 3.3
Figure 57 – Digital Pots Menu Screen
Displays the gain of the PGA.
Reduces the raw signal to measurable level.
The concentration voltage measured by the analog to
digital converter.
The concentration voltage after adjustment for the
PGA gain factor.
Electronic zero for the measure channel.
Electronic zero for the measure channel.
The reference voltage of the detector.
Adjusts the drive current for the IR source.
The voltage driving the IR source.
Adjusts the drive current for the cooler.
The voltage being applied to the cooler.
This Digital Pot is used for diagnostics only. When in
the Electrical or Preamp Diagnostic Mode, this
Digital pot should be adjusted from 0 - 255. This will
produce a change in the concentration voltage as
well as the indicated gas concentration.
This Digital Pot is used for diagnostics only. When in
the Electrical or Preamp Diagnostic Mode, this
Digital Pot should be adjusted from 0 - 255. This will
produce a change in the reference voltage.
Operate (default): Puts the instrument in normal
operation mode.
Electrical: Injects an artificial test signal into the
electronic processing circuitry on the main controller
PCA to verify that the circuitry is operating correctly.
When in this Diagnostic Mode, adjust the Test Meas
Pot from 0 - 255. This will produce a change in the
concentration voltage as well as the indicated gas
concentration. Adjust the Test Ref Pot from 0 - 255
to produce a change in the reference voltage.