EasyManua.ls Logo

Ecotech Acoem Serinus 30 - 3. Operation; 3.1 Warm-Up; 3.2 Measurement

Ecotech Acoem Serinus 30
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
3. Operation
3.1 Warm-Up
When the instrument is first turned ON it must go through a period of adjustment and calibration. No
measurements are taken during this warm-up period.
The following activities occur during warm-up:
Auto Ref. Adjust
Adjusts the reference voltage and internal trim pot setting
Auto Zero Adjust
Adjusts the zero offset
Heat CO Scrubber
The instrument progressively increases the temperature of the CO scrubber until it reaches the desired
temperature of 90 °C. When heating from cold, the scrubber will take approximately 20 - 30 minutes
to reach desired temperature.
Cell Temperature
Waits until the cell temperature reaches at least 90% of set point.
After this warm-up has completed the instrument will immediately begin making measurements (refer
to Section 3.2).
3.2 Measurement
The Serinus 30 primarily operates in the sample cycle continuously. A background is performed after
warm-up, once a day (at 23:45 as default) or when the chassis temperature changes by 4 °C. It is used
to measure background absorption in the cell and is subtracted from sample measurements.
Table 4 Measurements: Background and Sample Cycles
Instrument State
Duration (minutes)
Description
Background Fill
3
Measurement cell fills with
background air
Background Meas
1
Measurement of background air
CO Sample Fill
3
Measurement cell fills with sample air
CO Sample Meas
Continuous
Measurement of sample air

Table of Contents

Related product manuals