Displays the interval at which the data is saved to the USB
memory stick. Selecting a 1 sec interval may result in occasional
measurements not being logged or slow response to serial
commands.
When enabled displays the interval at which the data is saved to
the USB memory stick for the calibration time period. This
occurs when the system is in Cal. Mode SPAN or ZERO.
Selecting a 1 sec interval may result in occasional
measurements not being logged or slow response to serial
commands. Seconds (1, 2, 5, 10, 15, 30) or Minutes (1, 2, 3, 5,
10, 15, 20 or 30) or hours (1, 2, 4, 6, 8, 12 or 24). Default is
Disabled.
Numeric list of the parameters logged. This is a quicker way to
enter parameters (for a list of parameters refer to Table 33).
Select the list of logged parameters by name.
3.4.24 Serial Communication Menu
Main Menu → Communications Menu → Serial Communication Menu
Figure 66 – Serial Communication Menu Screen
This is the ID of the instrument when using multidrop RS232
communications. This ID can be changed to support multiple
instruments on the same RS232 cable.
Bayern-Hessen ID
[Bayern-Hessen Protocol]
This is the Bayern-Hessen ID used by the Bayern-Hessen
Protocol.
CO ID
[Bayern-Hessen Protocol]
This is the ID of the CO gas used by the Bayern-Hessen
Protocol.
CO2 ID
[Bayern-Hessen Protocol]
[Input 1 is CO2 Option]
This is the ID of the CO
2
gas used by the Bayern-Hessen
Protocol. This option is only available if the user enables the
Input 1 is CO2 option in the Hardware Menu (refer to
Section 3.4.34).
Service port (RS232 #1)
Multidrop port (RS232 #2)
The port parameters below are repeated for each serial
port.