EasyManua.ls Logo

Ecotech Acoem Serinus 30 - Trace Level Instrument

Ecotech Acoem Serinus 30
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Serinus® 30 User Manual 3.3
Page 178
8.7 Trace Level Instrument (PN: E020127)
The trace option enables the instrument to detect levels carbon monoxide (CO) from 0 - 20,000 ppb
with a lower detectable limit of 20 ppb. This allows the Serinus 30 Trace to sample in the most useful
range for CO background measurement (25 - 500 ppb CO in air, levels that represent the range of global
CO mixing ratios in the unpolluted atmosphere).
The Serinus 30 Trace has a number of extra passes in the measurement cell to increase the resolution,
along with a thermally stabilised detector housing with a more sensitive IR detector operating at a
lower temperature (-50 °C) to reduce thermal noise.
This option is also available without a dryer (PN: E020120).
8.7.1 Trace Specifications
The following tables list the changes to a standard instrument with the trace option installed.
Table 13 Trace Parts Removed
Parts Removed
Quantity
Part Numbers
Measurement Cell
1
H014100
Table 14 Trace Parts Added
Parts Added
Quantity
Part Number
Trace Measurement Cell
1
H014100-02
Dryer
1
H014180
8.7.2 Measurement
Range
0 - 20 ppm (autoranging)
Lower detectable limit: 20 ppb, with Kalman filter active
8.7.3 Precision/Accuracy
Noise at Zero
< 10 ppb
8.7.4 Calibration
Zero Drift
24 hours: < 50 ppb