Serinus® 30 User Manual 3.3
Set to MANUAL to disable the automatic pump control. AUTO allows
the flow PID to modify the pump coarse and fine settings. START will
transition to AUTO after one second.
Internal pump speed control (Coarse). This field is only editable when
the Pump Control field is set to MANUAL.
Internal Pump speed control (Fine). This field is only editable when the
Pump Control field is set to MANUAL.
3.4.13 Service Menu
Main Menu → Service Menu
Figure 55 – Service Menu Screen
Loads the auto-backup configuration file. The configuration is
automatically backed up every night at midnight.
Loads a user selectable configuration file from the USB memory
stick.
Saves all of the EEPROM-stored user-selectable instrument
configurations to the USB memory stick (calibration and
communication settings, units, instrument gain, etc.). If there are
problems with the instrument use this function to save settings to
the USB memory stick and send this file (together with the
parameter list save) to your supplier with your service enquiry.
Saves a text file of various parameters and calculation factors. If
the user has problems with the instrument use this function to
save settings to the USB memory stick and send this file (together
with the configuration save) to your supplier with your service
enquiry.
This field allows the instrument to be set to either Online (normal
instrument operation) or In Maintenance (data is flagged as
invalid).
Displays when the next scheduled service is due.